Main feature
1. Be controlled by microprocessor
2. Gas driven electronically controlled, With high-intelligent control system. 
3. Used for adult and pediatric. 
4. Color TFT 12.1-inch screen display for ventilation parameters (Optional: Touch screen)
5. Humidifier with temperature adjustment
6. Flow (optional) and pressure trigger
7. PEEP function.
8. CPAP function. 
9. SIGH function. 
10. I/E function. 
11. Compliance compensated function. 
12. Inspiratory pressure and flow volume trigger function. 

Specification
Main technical parameter
Mode IPPV(MMV),IPPV+SIGH,A/C,SIMV,PCV,SPONT,MANU,CPAP, VCV, PEEP
Optional: PSV,BIPAP
Tidal volume 20-2000mL. 
Frequency 1-100/min 
Inhalation time 0~12s (I:E ration: 4:1~1:8)
Holding time 0~6s
Controlling pressure 0.2~6.0KPa 
PEEP/CPAP 0~4.0KPa
Inhalation trigger Pressure: -2.0~0kPa Flow: 1~30L/min
SIGH time 1-5 times (every 100 times)
Oxygen concentration 21%~100% (continual adjustable)
Ventilation parameter monitoring Tidal volume, minute ventilation, frequency, inhalation time, airway peak pressure, PEEP, airway mean pressure, O2 concentration, Time-airway pressure waveform, Time- flow waveform, and trigger indicate.
Alarm: ( Audio-visual) No tidal volume (asphyxia), minute ventilation is too high (low), airway pressure is too high (low), O2 concentration is too high (low), oxygen or air is not full, frequency is too high (under SPONT mode) AC power off, Back-up battery voltage is too low.
Power off There will be audio and visual alarm indicates when power is off in unusual way, and the power will be supplied by back-up battery( battery working time≥3 hours).
Gas resource oxygen or depurative air. The given working pressure is 280kPa~600kPa
Power ~220V 50Hz 
Air compressor specification
Output pressure 0.4MPa±0.1Mpa 
Continuous flow 45L/min
Noise ≤50dB(A)
Power AC 220V, 50Hz
Working environment  
Temperature +5ºC-+40ºC,Relative humidity: 20%-80%
Rating consume power ≤650W
Optional
Color TFT 12.1-inch touch screen 
Air compressor